Novel bone SPECT-CT methods – Optimization and validation for imaging bone metastases
Arvola, Samuli (2024-09-27)
Novel bone SPECT-CT methods – Optimization and validation for imaging bone metastases
Arvola, Samuli
(27.09.2024)
Turun yliopisto
Julkaisun pysyvä osoite on:
https://urn.fi/URN:ISBN:978-951-29-9840-1
https://urn.fi/URN:ISBN:978-951-29-9840-1
Tiivistelmä
Bone single-photon emission computed tomography (SPECT) combined with X-ray computed tomography (SPECT-CT) is expected to replace planar bone scintigraphy (BS) as the routine method for imaging bone metastases. In preparation for this change, this thesis aimed to optimize and validate novel SPECT-CT methods, including fast whole-body image acquisition using a cadmium-zinc-telluride (CZT) SPECT-CT system, reprojected bone SPECT-CT (RBS), and quantitative SPECT.
Fast image acquisition was optimized by comparing the sensitivity and specificity between SPECT images acquired from the same patients with 50-, 32-, 26-, and 16-min acquisition times. RBS was validated by comparing the sensitivity and specificity between BS and RBS images of the same patients. Quantitative SPECT was validated by comparing standardized uptake values (SUVs) of the same bone metastases between SPECT and positron emission tomography (PET).
The average patient-level sensitivities for the 50-, 32-, 26-, and 16-min images were 88, 92, 100, and 96%, respectively, and the corresponding specificities 78, 84, 84, and 78%, respectively. The average patient-level sensitivities for BS and RBS were 75 and 87%, respectively, and the corresponding specificities 79 and 39%, respectively. SUVs correlated strongly (R2 ≥ 0.80) between SPECT and PET.
Whole-body bone SPECT-CT can be performed using a CZT system in less than 20 min without loss of diagnostic performance. Whole-body bone SPECT-CT can be reprojected into planar images with excellent sensitivity but limited specificity for identifying bone metastases. The strong correlation of SUVs between SPECT and PET demonstrates that SPECT SUVs are feasible for uptake measurements in bone metastases. Uudet luuston SPECT-TT-menetelmät — Optimointi ja validointi luustoetäpesäkkeiden kuvantamiseen
Luuston yksifotoniemissiotomografian (SPECT) yhdistettynä tietokonetomografiaan (SPECT-TT) odotetaan syrjäyttävän tasokuvauksen rutiinimenetelmänä luustoetäpesäkkeiden kuvantamisessa. Tämän muutoksen innoittamana tämä väitöskirja pyrki optimoimaan ja validoimaan uusia SPECT-TT-menetelmiä, kuten nopea koko kehon kuvaus kadmium-sinkki-telluuri-SPECT-TT-laitteella, SPECT-TT kuvista lasketut tasokuvat ja kvantitatiivinen SPECT.
Nopea kuvaus optimoitiin vertaamalla herkkyyttä ja spesifisyyttä samoista potilaista 50, 32, 26 ja 16 min kuvausajoilla otettujen SPECT-kuvien välillä. Lasketut tasokuvat validoitiin vertaamalla herkkyyttä ja spesifisyyttä samojen potilaiden kuvattujen ja laskettujen tasokuvien välillä. Kvantitatiivinen SPECT validoitiin vertaamalla samojen luustoetäpesäkkeiden standardoituja kertymäarvoja SPECT- ja positroniemissiotomografia (PET) -kuvien välillä.
Keskimääräiset potilastason herkkyydet 50, 32, 26 ja 16 min kuvilla olivat 88, 92, 100 ja 96 % ja spesifisyydet 78, 84, 84 ja 78 %. Keskimääräiset potilastason herkkyydet kuvatuilla ja lasketuilla tasokuvilla olivat 75 ja 87 % ja spesifisyydet 79 ja 39 %. Standardoidut kertymäarvot korreloivat vahvasti (R2 ≥ 0.80) SPECT- ja PET-kuvien välillä.
Koko kehon luuston SPECT-TT voidaan kuvata kadmium-sinkki-telluuri-laitteella alle 20 minuutissa ilman diagnostisen suorituskyvyn menetystä. Koko kehon SPECT-TT-kuvasta voidaan laskea tasokuvat, joilla on erinomainen herkkyys mutta rajoitettu spesifisyys luustoetäpesäkkeiden tunnistamisessa. Vahva korrelaatio SPECT- ja PET-kuvien standardoitujen kertymäarvojen välillä osoittaa, että SPECT-kuvien standardoidut kertymäarvot soveltuvat kertymämittauksiin luustoetäpesäkkeissä.
Fast image acquisition was optimized by comparing the sensitivity and specificity between SPECT images acquired from the same patients with 50-, 32-, 26-, and 16-min acquisition times. RBS was validated by comparing the sensitivity and specificity between BS and RBS images of the same patients. Quantitative SPECT was validated by comparing standardized uptake values (SUVs) of the same bone metastases between SPECT and positron emission tomography (PET).
The average patient-level sensitivities for the 50-, 32-, 26-, and 16-min images were 88, 92, 100, and 96%, respectively, and the corresponding specificities 78, 84, 84, and 78%, respectively. The average patient-level sensitivities for BS and RBS were 75 and 87%, respectively, and the corresponding specificities 79 and 39%, respectively. SUVs correlated strongly (R2 ≥ 0.80) between SPECT and PET.
Whole-body bone SPECT-CT can be performed using a CZT system in less than 20 min without loss of diagnostic performance. Whole-body bone SPECT-CT can be reprojected into planar images with excellent sensitivity but limited specificity for identifying bone metastases. The strong correlation of SUVs between SPECT and PET demonstrates that SPECT SUVs are feasible for uptake measurements in bone metastases.
Luuston yksifotoniemissiotomografian (SPECT) yhdistettynä tietokonetomografiaan (SPECT-TT) odotetaan syrjäyttävän tasokuvauksen rutiinimenetelmänä luustoetäpesäkkeiden kuvantamisessa. Tämän muutoksen innoittamana tämä väitöskirja pyrki optimoimaan ja validoimaan uusia SPECT-TT-menetelmiä, kuten nopea koko kehon kuvaus kadmium-sinkki-telluuri-SPECT-TT-laitteella, SPECT-TT kuvista lasketut tasokuvat ja kvantitatiivinen SPECT.
Nopea kuvaus optimoitiin vertaamalla herkkyyttä ja spesifisyyttä samoista potilaista 50, 32, 26 ja 16 min kuvausajoilla otettujen SPECT-kuvien välillä. Lasketut tasokuvat validoitiin vertaamalla herkkyyttä ja spesifisyyttä samojen potilaiden kuvattujen ja laskettujen tasokuvien välillä. Kvantitatiivinen SPECT validoitiin vertaamalla samojen luustoetäpesäkkeiden standardoituja kertymäarvoja SPECT- ja positroniemissiotomografia (PET) -kuvien välillä.
Keskimääräiset potilastason herkkyydet 50, 32, 26 ja 16 min kuvilla olivat 88, 92, 100 ja 96 % ja spesifisyydet 78, 84, 84 ja 78 %. Keskimääräiset potilastason herkkyydet kuvatuilla ja lasketuilla tasokuvilla olivat 75 ja 87 % ja spesifisyydet 79 ja 39 %. Standardoidut kertymäarvot korreloivat vahvasti (R2 ≥ 0.80) SPECT- ja PET-kuvien välillä.
Koko kehon luuston SPECT-TT voidaan kuvata kadmium-sinkki-telluuri-laitteella alle 20 minuutissa ilman diagnostisen suorituskyvyn menetystä. Koko kehon SPECT-TT-kuvasta voidaan laskea tasokuvat, joilla on erinomainen herkkyys mutta rajoitettu spesifisyys luustoetäpesäkkeiden tunnistamisessa. Vahva korrelaatio SPECT- ja PET-kuvien standardoitujen kertymäarvojen välillä osoittaa, että SPECT-kuvien standardoidut kertymäarvot soveltuvat kertymämittauksiin luustoetäpesäkkeissä.
Kokoelmat
- Väitöskirjat [2825]